System Specifications
TABLE / TUBESTAND
Unique column and covers safeguard staff and patients from exposed electrical wires
Rounded corners prevent hip-level bruising common to veterinary technicians
Stationary table top: 56 in. x 24 in.
Welded base construction
Transformer mounting shelf
Grid cabinet with cassette tray
Grid cabinet / tubestand interlock
Grid 8:1 ratio, 103 LPI
Animal restraining device
Angulation dial and deluxe operator handgrips
Tubestand handles
COLLIMATOR
Manual collimator with LED light field and swivel mount Field light “on” time – selectable 30, 45 or 60 seconds
HIGH FREQUENCY GENERATOR
20 KHz High Frequency technology
30 kW rating
kVp range: 40 – 125, in 1 kV increments
mA stations: 50s, 100s, 100L, 300L
mAs range: 0.5 to 400
mAs/kVp technique selection
Digital kVp, mA, and mAs readout
Built-in tube protection and self diagnostics
Automatic generator “Shut Off” circuit
Adjustable volume, audible exposure signal
UL classified
X-RAY TUBE
1.0-2.0 mm F.S., 140,000 H.U. with 90° horns